NETZSCH High Temperature TGA
DESCRIPTION
The water vapor furnace with perfectly balanced accessories for steam generation,
gas mixing and flow control offers the unique opportunity to study the mass changes
(and energy changes with the STA 449 F3 Jupiter) in samples up to 1250°C
under defined absolute humidity.
Temperature range: RT to 1250°C
Heating and cooling rates: 0.001 K/min to 50 K/min
Weighing range: 35000 mg
TGA resolution: 0.1 µg (over entire weighing range)
Atmospheres: inert, oxidizing, reducing, static, dynamic
LOCATION
Lab EMC2 (1A First Floor)
CONTACT INFO
Dr. Thomas Delclos,
Bruker Vertex 80v FT-IR
DESCRIPTION
Fourier-Transform-Infrared (FT-IR) spectroscopy is an excellent solution for Material analysis or gas analysis.
IR spectroscopy is a classical technique, which depends upon the interaction of infrared radiation with the vibrating dipole moments of molecules.
ATR accessory:
Single reflection diamond crystal, all reflective, gold coated optics, no fragile composite material construction
Spectral range: 8,000 cm-1 - 10 cm-1
Exchangeable crystal plates
Ergonomic one-finger clamp mechanism with pressure control
Adjustable and reproducible sample pressure
Working distance (max. sample height): 20 mm
Rotatable pressure applicator moves out of the way, allowing easy cleaning
Easy exchange of QuickLock accessories with automatic recognition,
performance test and automatic load of measurement parameters
Purgeable and Fully integrated dust-proof cover
DRIFT accessory:
Praying Mantis from Harrick Scientific
LOCATION
EMC2 Lab 1A First Floor
CONTACT INFO
Dr. Thomas Delclos,
NETZSCH High temperature DSC
DESCRIPTION
High-Temperature Differential Scanning Calorimeter,
is specially tailored to the requirements of comparative material characterization
and quality control.
Platinum furnace,
forced air cooling system,
Temperature range: RT to 1500 °C,
Heating rates: 0.001 K/min to 50 K/min,
Measuring sensors for DSC and DTA,
Thermocouple types: S,
Atmospheres: inert, oxidizing, static, dynamic.
LOCATION
SAN Campus Arzanah 8023
CONTACT INFO
Dr. Thomas Delclos,
Nabertherm High temperature Furnace
DESCRIPTION
Ref:LHT 04/16
T max: 1600 C
Inner Dimension in mm (w, d, h): 150, 150, 150.
These furnaces were specially developed to determine combustion loss during annealing
and for thermogravimetric analysis (TGA) in the lab.
The complete system consists of the high-temperature furnace for 1600 °C, a table frame,
precision scale with feedthroughs into the furnace
and powerful software for recording both the temperature curve and the weight loss over time.
LOCATION
EMC2 Lab (1A, first Floor)
CONTACT INFO
Dr. Thomas Delclos,
DMA TA Q800
DESCRIPTION
TA Q800 is a state-of-the-art, non-contact,
linear drive technology to provide precise control
of stress, and air bearings for low friction support.
Strain is measured using optical encoder technology
that provides unmatched sensitivity and resolution.
Ideal for stiff samples and especially composite materials.
Specifications:
Maximum Force 18 N
Minimum Force 0.0001 N
Force Resolution 0.00001 N
Strain Resolution 1 nanometer
Modulus Range 10^3 to 3x10^12 Pa
Modulus Precision ± 1%
Tan A Sensitivity 0.0001
Tan A Resolution 0.00001
Frequency Range 0.01 to 200 Hz
Dynamic Sample Deformation Range ± 0.5 to 10 000 µm
Temperature Range -150 to 600 °C
Heating Rate 0.1 to 20 °C/min
Cooling Rate 0.1 to 10 °C/min
Isothermal Stability ± 0.1 °C
Time/Temperature Superposition Yes
LOCATION
Materials Testing Lab (undercroft)
CONTACT INFO
Dr. Thomas Delclos,